Pool Test 7 Professional Testing Kit for Pools and Hot Tubs
POOL-TEST-KIT
Measure all of the most important pool and hot tub parameters with Pool Test 7. This self-contained pool measuring kit includes all of the necessary instruments to test seven pool parameters accurately and reliably. The full testing kit, including accessories, is contained in a practical case for easy transport to different sites, making Pool Test 7 an ideal choice for pool mechanics and service technicians alongside businesses and private pool owners.
Pool Test 7 Professional Testing Kit for Pools and Hot Tubs
The Pool Test 7 Kit is a comprehensive testing solution for all important pool parameters. Four different instruments for testing are included alongside all of the required accessories for easy digital testing. This all-in-one pool testing kit provides a high-precision alternative to testing strips or chemical kits, with digital measurements displayed on the screen of each pool meter and no need for guesswork in determining the results of testing.
Want to speak to the experts? If you need professional guidance on how to test the water in swimming pools or you’d like to learn more about Hanna’s Pool Line range, contact our pool ambassadors directly today.
What’s Included?
Pool Test 7 includes testing instrumentation for seven of the most common parameters tested in hot tubs, swimming pools and spas. These parameters are covered with four different instruments:
HI-971044 Professional Photometer
The HI-971044 Pool Line Photometer is a multiparameter solution for accurate, digital measurement of alkalinity, free and total chlorine, cyanuric acid and pH. A fully waterproof, floating case and a complete on-screen tutorial make this advanced photometer suitable for personal and professional use. Learn more about this photometer here.
HI-7134 Phosphate Checker
The HI-7134 Low Range Phosphate Checker is a compact digital meter for the reliable measurement of phosphate in pools and hot tubs, providing an ideal, accurate alternative to testing strips. Learn more about this digital pool tester here.
HI-981304 pH, EC and TDS Tester
The HI-981304 Handheld Pool Tester provides an easy-to-handle solution for testing pH, EC and TDS in pool water. Simply dip the tester in the pool or hot tub and take your reading from the clear, digital screen. Learn more about this multiparameter pool tester here.
HI-98319 Salinity Tester
The HI-98319 Pool Salinity Tester is an ideal way to precisely measure the salinity in a range of environments, including outdoor and indoor pools and hot tubs. Dip the tester directly in the water to get an accurate measurement in parts per thousand (ppt), Practical Salinity Units (PSU), or Specific Gravity (SG). Learn more about this salt pool tester here.
In addition to four of Hanna’s exceptionally accurate digital meters for pools, the Pool Test 7 also includes all of the necessary accessories and equipment for testing directly out of the box. These include:
- HI-93701-T Free and Total Chlorine liquid reagent - 300 Tests
- HI-93722-01 Cyanuric Acid reagent - 100 Tests
- HI-775-26 Reagents for freshwater Alkalinity Checker - 25 tests
- HI-70300s pH electrode storage solution - 25 ml
- HI-93755-53 Chlorine Remover Reagent
- HI-740036P 100 ml beaker
- HI-731318 Glass cleaning cloth
- Bottle Brush, Scissors, 10 ml syringe & 1 ml syringe
Once you have used the initial reagent packaged with Pool Test 7, these chemicals can be easily ordered online with free next-day delivery on all orders over £50.
Pool Test 7 features
Comprehensive coverage
Pool Test 7 covers all of the parameters you need to measure water in swimming pools and hot tubs in one portable kit. This complete coverage ensures exceptional accuracy in all pool testings, from pH to chlorine to cyanuric acid. Everything is included in the kit for you to start testing immediately, providing the information you need to effectively dose the water and make it safe for swimmers and bathers.
Convenience in a single kit
This all-in-one kit includes a specially designed hardwearing case with a foam insert, ensuring all meters and accessories are secure and safe during travel. The portable nature of Pool Test 7 makes this kit easy to transport, whether you need to measure multiple facilities, store the equipment safely nearby or travel to multiple sites. Each tester and meter is handheld and easy to use beside the pool or hot tub for your convenience.
Fast, easy digital testing
Hanna’s digital testing instrumentation for pools is exceptionally accurate, providing a convenient and reliable alternative to chemical test kits or pool testing strips. Digital testing provides precision in measurement, providing the information needed to correctly manage and dose the water in pools and hot tubs. This helps to prevent safety issues and ensures the water is sanitised and meets the necessary standards, such as HSG282 for businesses.
FAQs
What is included in the Pool Test 7 kit?
The entire contents of the Pool Test 7 pool water testing kit include:
- HI-971044 Photometer for Alkalinity, Free and Total Chlorine, Cyanuric Acid and pH
- HI-7134 Phosphate Checker and reagents for 25 tests
- HI-981304 Digital pH, EC & TDS Tester
- HI-98319 Digital Salinity Tester
- HI-93701-T Free and Total Chlorine liquid reagent - 300 Tests
- HI-93722-01 Cyanuric Acid reagent - 100 Tests
- HI-775-26 Reagents for freshwater Alkalinity Checker - 25 tests
- HI-70300s pH electrode storage solution - 25 ml
- HI-93755-53 Chlorine Remover Reagent
- HI-740036P 100 ml beaker
- HI-731318 Glass cleaning cloth
- Bottle Brush, Scissors, 10 ml syringe & 1 ml syringe
This kit is packaged safely into a purpose-designed hard case with foam inserts, preserving the kit for day-to-day use during storage and transportation.
Why should I buy a complete pool testing kit?
A complete kit for testing swimming pools and hot tubs is an ideal solution if you’d like to upgrade from test strips or chemical test kits. This kit includes all of the necessary meters to test immediately out of the box, making it a suitable option for anyone purchasing pool testing equipment for the first time as a private or professional owner. The exceptional accuracy of Hanna testing instrumentation makes Pool Kit 7 the best pool testing kit for a range of different requirements and safety standards.
Can I get a demonstration of Pool Line products?
Our pool ambassadors are on hand to provide support and guidance on our full range of pool and hut tub testers and meters. Get in contact with them directly at 01525 850 855 or email us at sales@hannainstruments.co.uk to discuss your specific requirements for pool water measurement.
Where can I buy Pool Test 7?
The Pool Test 7 photometer kit is available for purchase online today from Hanna Instruments UK. Our range of reagents is also readily available for purchase, with stock held in the UK for next-day delivery on all key solutions.
Can I get advice on the best pool testers for my requirements?
Speak to our pool ambassadors directly to chat with the experts about the best hot tub testers and pool meters to meet your requirements. Our Pool Line range includes a wide assortment of versatile testers, photometers and fully automatic pool controllers to suit your needs. Call 01525 850 855 or email us at sales@hannainstruments.co.uk today.
Read More
HI-971044 Specifications | ||
pH | ||
Range | 6.5 to 8.5 | |
Resolution | 0.1 pH | |
Accuracy @25°C | ±0.1 pH of reading | |
Method | adaptation of the Phenol Red method | |
Alkalinity | ||
Range | 0 to 500 mg/L (as CaCO3) | |
Resolution | 1 mg/L | |
Accuracy @ 25°C | ±5 mg/L ±5% of reading | |
Method | Colorimetric method | |
Free and Total Chlorine | ||
Range (all methods) | 0.00 to 5.00 mg/L (as Cl2) | |
Resolution | 0.01 mg/L | |
Accuracy @ 25°C | ±0.03 mg/L ±3% of reading | |
Method | adaptation of the EPA DPD method 330.5 | |
Cyanuric Acid | ||
Range | 0 to 80 mg/L (as CYA) | |
Resolution | 1 mg/L | |
Accuracy @ 25°C | ±1 mg/L ±15% of reading | |
Method | adaptation of the turbidimetric method | |
Measurement System | ||
Light source | light emitting diode | |
Bandpass filters | 525 nm and 610 nm | |
Bandpass filters bandwidth | 8nm | |
Bandpass filter wavelength accuracy | ±1.0nm | |
Light detector | silicon photocell | |
Additional specifications | ||
Auto logging | 50 readings | |
Display | 128 x 64 pixel LCD with backlight | |
Auto-off | after 15 minutes of inactivity | |
Battery type and life | alkaline 1.5 AA x 3 / >800 measurements without backlight | |
Environment | 0 to 50°C; 0 to 100% RH, | |
Dimensions | 142.5 x 102.5 x 50.5 mm | |
Weight | 380g | |
IP rating | IP67 |
HI-7134 Specifications | ||
Range | 0.00 to 2.50 ppm (mg/L) | |
Resolution | 0.01 ppm (mg/L) | |
Precision | ±0.04 ppm (mg/L) ±4% of reading @ 25°C | |
Light Source | LED @ 525 nm | |
Light Detector | Silicon photocell | |
Environment | 0 to 50°C (32 to 122°F); RH max 95% non-condensing | |
Battery Type | (1) 1.5V AAA | |
Auto-off | After 10 minutes of non-use and after read | |
Dimensions | 86 x 61 x 37.5 mm (3.2 x 2.4 x 1.5”) | |
Weight | 64 g (2.25 oz.) | |
Method | Adaptation of Standard Methods for the Examination of water and wastewater. 20th edition, Ascorbic Acid method. The reaction between phosphate and the reagent causes a blue tint in the sample. | |
PLEASE NOTE: | In some cases, samples may contain other parameters which may interfere with results | |
Interference may be caused by: | Iron above 50 mg/L Silica above 50 mg/L Silicate above 10 mg/L Copper above 10 mg/L Hydrogen sulphide, arsenate, turbid sample and highly buffered samples also interfere. |
HI-981304 Specifications | ||
pH Range | 0.00 to 14.00 | |
pH Resolution | 0.01 pH | |
pH Accuracy | ±0.05pH | |
Conductivity Range | 0.00 to 20.00 mS/cm | |
Conductivity Resolution | 0.01 mS/cm | |
Conductivity Accuracy | ±2% F.S. | |
TDS Range | 0.00 to 10.00 g/L (ppt) | |
TDS Resolution | 0.01 ppt (g/L) | |
TDS Accuracy | ±2% F.S. | |
Temperature Range | 0.0 to 60.0°C | |
Temp Resolution | 0.1°C | |
Temp Accuracy | ±0.5°C | |
Additional specifications | ||
EC/TDS calibration | Automatic, one point at 12.88 mS/cm or 6.44 ppt (g/L) | |
pH Calibration | Automatic, one or two point with two sets of standard buffers | |
Temp compensation | pH automatic; EC/TDS automatic with β adjustable from 0.0 to 2.4%/°C | |
TDS conversion factor | 0.45 to 1.00 | |
pH electrode replacement | HI-73127 | |
Battery life & type | 1.5V (4) approx. 300 hours of continuous use; auto off after 8 minutes of non-use | |
Environment | -5 to 50°C; RH max 100% | |
Dimensions | 163 x 40 x 26mm | |
Weight | 100 g |
HI-98319 Specifications | ||
Low Range: | ||
ppt (g/L) | Range: 0.00 to 10.00 ppt (g/L) | |
PSU | Range: 0.00 to 10.00 PSU | |
S.G. | Range: 1.000-1.007 | |
Resolution | ||
ppt (g/L) | 0.01 | |
PSU | 0.01 | |
SG | 0.001 | |
Accuracy | ||
ppt (g/L) | ±0.20 | |
PSU | ±0.20 | |
SG | ±0.001 |
High Range: | ||
ppt (g/L) | Range: 0.0 to 70.0 ppt (g/L) | |
PSU | Range: 0.0 to 70.0 PSU | |
S.G. | Range: 0.001-1.041 | |
Resolution | ||
ppt (g/L) | 0.1 | |
PSU | 0.1 | |
SG | 0.001 | |
Accuracy | ||
ppt (g/L) | ±1.0 (0.0 to 40.0) ±2.0 (40.0 to 70.0) | |
PSU | ±1.0 (0.0 to 40.0) ±2.0 (40.0 to 70.0) | |
SG | ±0.001 |
Automatic | ||
ppt (g/L) | Range: 0.00 to 9.99 ppt (g/L) 10.0 to 70.0 | |
PSU | Range: 0.00 to 9.99 ppt (g/L) 10.0 to 70.0 | |
S.G. | Range: 1.000 to 1.041 | |
Resolution | ||
ppt (g/L) | 0.01/0.1 | |
PSU | 0.01/0.1 | |
SG | 0.001 | |
Accuracy | ||
ppt (g/L) | ±0.20 (0.00 to 9.99) ±1.0 (10.0 to 40.0) ±2.0 (40.0 to 70.0) | |
PSU | ±0.20 (0.00 to 9.99) ±1.0 (10.0 to 40.0) ±2.0 (40.0 to 70.0) | |
SG | ±0.001 | |
Temperature | 0.0 to 50.0 °C/32.0 to 122.0°F | |
Resolution | 0.1°C | |
Accuracy | ±0.5 °C (±1.0 °F) | |
Method | ||
ppt | International Oceanographic Tables, 1966 | |
PSU | Standard Methods for the Examination of Water and Wastewater, 2520 B, Electrical Conductivity Method | |
S.G. | Standard Methods for the Examination of Water and Wastewater, 2520 C, Density Method | |
Calibration Solution | HI-70023 (5.00ppt); HI-70024 (35.00 ppt) | |
Calibration | Automatic, one or two-point calibration at 5.00 ppt or 35.00 ppt | |
Temperature Compensation | Automatic from 5.0 to 50.0 °C | |
Environment | 0 to 50°C; RH max 100% | |
Battery Type | CR2032 3V Lithium-ion (1) | |
Battery Life | Approximately 100 hours of continuous use | |
Weight | 68g without battery |