HI-97709

Measures the manganese content in water samples in the 0.0 to 20.0 mg/L (ppm).

Manganese

£507.50 (exc.VAT)
£609.00 (tax incl.)
Order now for 2 weeks delivery
Accessories
Reference
Price
Quantity
Full Description Specification Manuals

The HI-97709 is an auto-diagnostic portable photometer that has an advanced optical system that uses a Light Emitting Diode (LED) and a narrow band interference filter that allows for accurate and repeatable readings.

Each HI-97709 is delivered in a cardboard box and is supplied with:

  • Sample cuvette (2 pcs.)
  • Sample cuvette cap (2 pcs.)
  • Plastic stopper (2 pcs.)
  • 1.5V AA Alkaline batteries
  • Instrument quality certificate
  • Instruction manual

About the Photometer

The optical system is sealed from outside dust, dirt and water. The meter uses an exclusive positive-locking system to ensure that the cuvettes are placed into the holder in the same position every time.

With the CAL Check™ functionality, users are able to validate the performance of the instrument at any time and apply a user calibration (if necessary). Hanna Instruments CAL Check cuvettes are made with NIST traceable standards.

The built-in tutorial mode guides users step-by-step through the measurement process. It includes all steps required for sample preparation, the required reagents and quantities.

The HI-97709 meter measures the manganese content in water samples in the 0.0 to 20.0 mg/L (ppm). The method is an adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th Edition, Periodate Method.

The HI-97709 photometer is a compact and versatile meter suitable for field or bench measurements, featuring a:

  • Sophisticated optical system
  • Meter validation using certified CAL Check cuvettes
  • Tutorial mode guides the user step-by-step
  • Auto logging
  • Waterproof IP67, floating case
  • GLP features

Optical system

The internal reference system (reference detector) of the HI-97709 photometer compensates for any drifts due to power fluctuations or ambient temperature changes, providing a stable source of light for your blank (zero) measurement and sample measurement.

LED light sources offer superior performance compared to tungsten lamps. LEDs have a much higher luminous efficiency, providing more light while using less power. They also produce little heat, which could otherwise affect electronic stability. LEDs are available in a wide array of wavelengths, whereas tungsten lamps have poor blue/violet light output.

Improved optical filters ensure greater wavelength accuracy and allow a brighter, stronger signal to be received. The end result is higher measurement stability and less wavelength error.

A focusing lens collects all of the light that exits the cuvette, eliminating errors from cuvette imperfections and scratches, eliminating the need to index the cuvette.

Reagent sets (buy separately)

  • HI-93709-01 Manganese HR Reagent - 100 tests
  • HI-93709-03 Manganese HR Reagent - 300 tests

Accessories

  • HI-7101412 blue carrying case for HI977xx and 2 CAL Check cuvettes
  • HI-731318 cloth for wiping cuvettes (4 pcs.)
  • HI-731331 glass cuvette (4 pcs.)
  • HI-731336N cap for glass cuvette (4 pcs.)
  • HI-93703-50 cuvette cleaning solution (230 mL)
  • HI-97709-11 CAL Check® standards for Manganese HR - cuvette kit

Read More

Range0.0 to 20.0 mg/L (as Mn)
Resolution0.1 mg/L
Accuracy±0.2 mg/L ±3% of reading
MethodAdaptation of Standard Methods for the Examination of Water and Wastewater, 18th Edition, Periodate Method
Photometer/Colorimeter Light SourceLight emitting diode
Photometer/Colorimeter Light DetectorSilicon photocell with narrow band interference filter @ 525 nm
Bandpass Filter Bandwidth8 nm
Bandpass Filter Wavelength Accuracy±1.0 nm
Cuvette Typeround, 24.6 mm diameter (22 mm inside)
GLPDate, time of last calibration
Display128 x 64 pixel B/W LCD with backlight
Logging Typeautomatic logging
Logging Memory50 measurement
Battery Type/Life1.5V AA battery (3 pcs.) / > 800 measurements (without backlight)
Automatic Shut-OffAfter 15 minutes of inactivity (30 minutes before a READ measurement)
Environment0 to 50°C (32 to 122°F); relative humidity: 100% (IP 67, floating case)
Weight380g with batteries
Dimensions142.5 x 102.5 x 50.5 mm
Ordering InformationHI-97709 is supplied with cuvettes (2), caps (2), plastic inserts for cuvettes (2), 1.5V AA batteries (3), instruction manual, and instrument quality certificate.