HI-83306-02 Multi-parameter Environmental Analysis Photometer with pH meter
In Stock 4 Items available
HI-83306-02
One meter can be used for both photometric and pH measurements. HI-83306 is a compact, multi-parameter photometer for use in the environmental lab or in the field.
HI-83306-02 Environmental Analysis Photometer with pH digital input
Advanced optical system - Unparalleled performance from a benchtop photometer
One meter can be used for both photometric and pH measurements. HI-83306 is a compact, multi-parameter photometer for use in the environmental lab or in the field. The meter is one of the most advanced photometers available with an innovative optical design that utilises a reference detector and focusing lens to eliminate errors from changes in the light source and from imperfections in the glass cuvette. This meter has 23 different programmed methods measuring 16 key water quality parameters and also offers an absorbance measurement mode for performance verification and for users who would like to develop their own concentration versus absorbance curves. Environmental parameters include nitrate, phosphorous and dissolved oxygen.
The photometer features an innovative optical system that uses LEDs, narrow band interference filters, a focusing lens, and a silicon photodetector for absorbance measurement and a reference detector to maintain a consistent light source which ensures accurate and repeatable photometric readings every time.
The HI-83306 was developed to measure the most common parameters in environmental water quality monitoring. Nutrients such as nitrates and phosphates are key indicators of nutrient pollution from agricultural sources and are considered dangerous to environmental waters. Too few nutrients and waters will be unable to sustain healthy ecosystems; too many nutrients and algal blooms can form, which can be detrimental to water quality and aquatic health. Dissolved oxygen is an essential to performing biological processes for many forms of aquatic life, such as fish, plants and microorganisms.
A digital pH electrode input allows the user to measure pH by a traditional glass electrode. The digital pH electrode has a built in microchip within the probe that stores all of the calibration information. Having the calibration information stored in the probe allows for hot swapping of pH electrodes without having to recalibrate. All pH measurements are automatically compensated for temperature variations with a built in thermistor located in the tip of the sensing bulb for fast and accurate temperature measurement.
The HI-83306 offers an absorbance measuring mode that allows for CAL Check standards to be used to validate the performance of the system. The absorbance mode allows the user to select one of the four wavelengths of light (420 nm, 525 nm, 575 nm, and 610 nm) to measure and plot their own concentration versus absorbance mode. This is useful for users with their own chemical method and for educators to teach the concept of absorbance by using the Beer-Lambert Law.
Two USB ports are provided for transferring data to a flash drive or computer and to use as a power source for the meter. For added convenience and portability, the meter can also operate on an internal 3.7 VDC Lithium-polymer rechargeable battery.
Key features:
- Backlit 128 x 64 Pixel Graphic Liquid Crystal Display
- Built-in Reaction Timer for Photometric Measurements
- Absorbance mode - concentration versus absorbance
- Units of Measurement plus chemical form displayed
- Result conversion at the touch of a button
- Cuvette cover
- pH and temperature measurement with a single probe
- Good Laboratory Practice (GLP) - track calibration information including date, time, buffers used, offset and slope for traceability
- pH CAL Check alerts user to potential problems during the calibration process
- Data Logging - Up to 1000 photometric and pH readings can be stored.
- Logged readings can be quickly and easily transferred to a flash drive or to a PC. Data is exported as a .CSV file for use with spreadsheet programs.
- Battery Status Indicator
- Error Messages
High Efficiency LED Light Source
An LED light source offers superior performance compared to a tungsten lamp. LEDs have a much higher luminous efficiency, providing more light while using less power. They also produce very little heat, which could otherwise affect the optical components and electronic stability. LEDs are available in a wide array of wavelengths, whereas tungsten lamps are supposed to be white light (all wavelengths of visible light) but actually have a poor blue/violet light output.
High Quality Narrow Band Interference Filters
The narrow band interference filter not only ensure greater wavelength accuracy (+/- 1 nm) but are extremely efficient. The filters used allow up to 95% of the light from the LED to be transmitted as compared to other filters that are only 75% efficient. The higher efficiency allows for a brighter, stronger light source. The end result is higher measurement stability and less wavelength error.
Reference Detector for a Stable Light Source
A beam splitter is used as part of the internal reference system of the HI-83300 photometer. The reference detector compensates for any drift due to power fluctuations or ambient temperature changes. Now you can rely on a stable source of light between your blank (zero) measurement and sample measurement.
Large Cuvette Size
The sample cell of the HI-83308 fits a round, glass cuvette with a 25 mm path length. Along with the advanced optical components, the larger size of the cuvette greatly reduces errors in rotation from the indexing mark of the cuvettes. The relatively long path length of the sample cuvette allows the light to pass through more of the sample solution, ensuring accurate measurements even in low absorbance samples.
Focusing Lens for Greater Light Yield
Adding a focusing lens to the optical path allows for the collection of all of the light that exits the cuvette and focusing the light on the silicon photo detector. This novel approach to photometric measurements cancels the errors from imperfections and scratches present in the glass cuvette eliminating the need to index the cuvette.
Read More
Input Channels | 1 pH electrode input and 5 photometer wavelengths | |
pH Electrode | Digital pH electrode (not included) | |
Logging type | Log on demand with user name and sample ID optional input | |
Logging Memory | 1000 readings | |
Connectivity | USB-A host for flash drive; micro-USB-B for power and computer connectivity | |
GLP | Calibration data for connected pH electrode | |
Display | 128 x 64 pixel LCD with backlight | |
Battery Type/Life | 3.7 VDC Li-polymer rechargeable battery/ >500 photometric measurements or 50 hours of continuous pH measurement | |
Power supply | 5 VDC USB 2.0 power adapter with USB-A to micro-USB-B cable (included) | |
Environment | 0 to 50.0°C (32 to 122.0°F; 0 to 95% RH, non-condensing | |
Dimensions | 206 x 177 x 97mm (8.1 x 7.0 x 3.8") | |
Weight | 1.0 kg (2.2lbs) | |
Photometer/Colorimeter Light Source | 5 LEDs with 420nm, 466nm, 525nm, 575nm, and 610nm narrow band interference filters | |
Photometer/Colorimeter Light Detector | Silicon Photodetector | |
Bandpass filter Bandwidth | 8nm | |
Bandpass filter wavelength accruacy | +-1nm | |
Cuvette Type | Round, 24.6mm | |
Number of Methods | 128 Max | |
Ordering Information | HI-83306 is supplied with sample cuvettes and caps (x4), cloth for wiping cuvettes, USB to micro USB cable connector, power adapter and instruction manual. |
AMMONIA
Ammonia Range | Low Range: 0.00 to 3.00 mg/L (as NH3-N) Medium Range: 0.00 to 10.00 mg/L (as NH3-N) High Range: 0.0 to 100.0 mg/L (as NH3-N) | |
Ammonia Resolution | 0.01 mg/L; 0.1 mg/L | |
Ammonia Accuracy | Low Range: ±0.04 mg/L ±4% of reading Medium Range: ±0.05 mg/L ±5% of reading High range: ±0.5 mg/L ±5% of reading at 25 °C | |
Ammonia Method | Adaptation of the ASTM Manual of Water and Environmental Technology, D1426-92, Nessler method |
CHLORINE
Free Chlorine Range | 0.00 to 5.00 mg/L (as Cl2) | |
Free Chlorine Resolution | 0.01 mg/L | |
Free Chlorine Accuracy | ±0.03 mg/L ±3% of reading | |
Total Chlorine Range | 0.00 to 5.00 mg/L (as Cl2) | |
Total Chlorine Resolution | 0.01 mg/L | |
Total Chlorine Accuracy | ±0.03 mg/L ±3% of reading | |
Chlorine Method | Adaptation of the EPA 330.5 DPD method |
CHROMIUM VI
Chromium VI Range | Low Range: 0 to 300 μg/L (as Cr6+) High Range: 0 to 1000 μg/L (as Cr6+) | |
Chromium VI Resolution | 1 μg/L | |
Chromium VI Accuracy | Low Range: ±10 μg/L ±4% of reading High Range: ±5 μg/L ±4% of reading | |
Chromium VI Method | Adaptation of the ASTM Manual of Water and Environmental Technology, D1687-92, Diphenylcarbohydrazide method. |
COLOUR, WATER
Colour, Water Range | 0 to 500 PCU (Platinum Cobalt Units) | |
Colour, Water Resolution | 1 PCU | |
Colour, Water Accuracy | ±10 PCU ±5% of reading | |
Colour, Water Method | Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, Colorimetric Platinum Cobalt method. |
COPPER
Copper Range | Low Range: 0.000 to 1.500 mg/L (as Cu2+) High range: 0.00 to 5.00 mg/L (as Cu2+) | |
Copper Resolution | 0.001 mg/L; 0.01 mg/L | |
Copper Accuracy | Low Range: ±0.010 mg/L ±5% of reading High Range ±0.02 mg/L ±4% of reading | |
Copper Method | Adaptation of the EPA method |
CYANURIC ACID
Cyanuric Acid Range | 0 to 80 mg/L (as CYA) | |
Cyanuric Acid Resolution | 1 mg/L | |
Cyanuric Acid Accuracy | ±1 mg/L ±15% of reading | |
Cyanuric Acid Method | Adaptation of the turbidimetric method |
MOLYBDENUM
Molybdenum Range | 0.0 to 40.0 mg/L (as Mo6+) | |
Molybdenum Resolution | 0.1 mg/L | |
Molybdenum Accuracy | ±0.3 mg/L ±5% of reading | |
Molybdenum Method | Adaptation of the mercaptoacetic acid method |
NICKEL
Nickel Range | Low Range: 0.000 to 1.000 mg/L (as Ni) High Range: 0.00 to 7.00 g/L (as Ni) | |
Nickel Resolution | 0.001 mg/L; 0.01 g/L | |
Nickel Accuracy | Low range: ±0.010 mg/L ±7% of reading High Range: ±0.07 g/L ±4% of reading | |
Nickel Method | Low Range: Adaptation of the PAN method High Range: Adaptation of the photometric method |
NITRATE
Nitrate Range | 0.0 to 30.0 mg/L (as NO3-- N) | |
Nitrate Resolution | 0.1 mg/L | |
Nitrate Accuracy | ±0.5 mg/L ±10% of reading | |
Nitrate Method | Cadmium reduction method | |
Reagent code | HI-93728-01 |
NITRITE
Nitrite Range | Freshwater High Range: 0 to 150 mg/L (as NO2-) | |
Nitrite Resolution | Freshwater: 1 mg/L | |
Nitrite Accuracy | Freshwater High Range: ±4 mg/L ±4% of reading | |
Nitrite Method | HR - Ferrous Sulphate |
OXYGEN, DISSOLVED
Oxygen, Dissolved Range | 0.0 to 10.0 mg/L (as O2) | |
Oxygen, Dissolved Resolution | 0.1 mg/L | |
Oxygen, Dissolved Accuracy | ±0.4 mg/L ±3% of reading | |
Oxygen, Dissolved Method | Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, Azide modified Winkler method |
pH
pH range | 6.5 to 8.5 pH | |
pH Resolution | 0.1 pH | |
pH Accuracy | ±0.1 pH | |
pH Method | Phenol red method |
PHOSPHATE
Phosphate Range | Low Range: 0.00 to 2.50 mg/L (as PO43-) High range: 0.0 to 30.0 mg/L (as PO43-) | |
Phosphate Resolution | 0.01 mg/L; 0.1 mg/L | |
Phosphate Accuracy | Low Range: ±0.04 mg/L ±4% of reading High Range: ±1 mg/L ±4% of reading | |
Phosphate Method | Low Range: Adaptation of the Ascorbic Acid method High Range: Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, Amino Acid method |
SILICA
Silica Range | Low Range: 0.00 to 2.00 mg/L (as SiO2) | |
Silica Resolution | 0.01 mg/L | |
Silica Accuracy | Low Range: ±0.03 mg/L ±3% of reading | |
Silica Method | Low Range: Adaptation of the ASTM Manual of Water and Environmental Technology, D859, Heteropoly Molybdenum Blue method |
SILVER
Silver Range | 0.000 to 1.000 mg/L (as Ag) | |
Silver Resolution | 0.001 mg/L | |
Silver Accuracy | ±0.020 mg/L ±5% of reading | |
Silver Method | Adaptation of the PAN method |
ZINC
Zinc Range | 0.00 to 3.00 mg/L (as Zn) | |
Zinc Resolution | 0.01 mg/L | |
Zinc Accuracy | ±0.03 mg/L ±3% of reading | |
Zinc Method | Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, Zincon method |