+44 (0)1525 850 855
Order before 4pm for same day dispatch Weekdays only
Menu
Hanna HI-50021P Check Solution, 20mL, 25 sachets (HI-9813-6 only)

In Stock 2 Items available

HI-50021P

HI-50021P is a premium quality calibration check solution used exclusively for the HI-9813-6 portable pH/EC/TDS/Temperature meter. 

25 x 20ml sachets

£47.60 (exc.VAT)
£57.12 (tax incl.)
Accessories
Reference
Price
Quantity
Full Description Specification

Each sachet has the lot number and expiry date stamped on it and is made of light block foil ensuring freshness each time one is opened.

The HI-50021P has 25 single use sachets containing 20 mL of solution that is used to verify the calibration and functionality of the pH electrode.

HI-50021P is a premium quality calibration check solution produced in accordance with ISO 3696/BS3978 standards using high purity salts, deionized water, certified weight-checked balances, and Class A glassware in a temperature controlled environment monitored with certified thermometers. It is specially formulated for use with the HI-9813-6 portable pH/EC/TDS/Temperature meter. It allows the user to check the meter’s calibration status at any time by simply rinsing the probe with water and immersing it in the solution. The meter will indicate the probe’s condition.

Convenience of individual packaging

  • Ideal for field use due to portability
  • Every sachet is as fresh as the day it was packaged

Each sachet marked with lot number and expiry date

  • Hanna solutions are specially formulated to have an expiry of 5 years from manufacture for an unopened sachet.

Light block packaging

  • Prevents any oxidation from UV light that could affect quality

CAL-CHECK™ FEATURE / CLEAN & CAL PROCEDURE

The unique Cal-Check™ feature allows the user to check the meter calibration status at any time. Simply follow the procedure outlined below for a quick assessment of the calibration status.

1) Rinse the probe with water.

2) Immerse the probe in the HI-50021 check solution.

3) Press the Check key.

4) If the meter is calibrated “Probe is OK” message is displayed.

5) If cleaning is needed “Clean Probe and Calibrate” message is displayed.

6) Immerse the probe in appropriate cleaning solution (i.e. HI-700661) for 5 minutes, then rinse it with tap water.

7) Immerse again in the HI-50021 check solution and press the CHECK key. If the message “Probe is OK” is displayed, the meter does not need calibration. If the message “Clean Probe and Calibrate” is displayed then calibration must be performed.

Read More

PackageSachets
Size20ml
Quality25
Certificate of AnalysisNo