HI-1002/5 Process PH Electrode
HI-1002/5
Industrial probe with a glass bodied electrode for use in aggressive chemicals. It is also easy to clean. A PEI protective sleeve gives the electrode resistance against mechanical stress.
Operating limits are -5 to 80°C (23 to 176°F) and pressure up to 6 bar (87 psi).
- ½” NPT external thread for in-line installation
- pH electrode with exclusive PTFE non-clogging membrane
- Perfect for continuous flow-thru monitoring in industrial water softening and cooling tower applications
In order to reduce normal contamination coming from industrial use, this electrode combines a polymer reference and double junction technology. With this technology, no refilling is required and the electrode can be used in samples such as organic compounds, proteins and heavy metals. In addition, the pH electrode uses a unique annular PTFE junction that minimises clogging.
Some models also feature a built-in amplifier, which allows for measurements to be taken far from the location of the instrument without requiring a transmitter. HI-1002 probes incorporate a BNC connector that enables connection to any pH meter quickly and easily. Models with 3 or 5 meter cable are available.
Read More
HI-1002/5 | ||
Junction | Double, PTFE | |
Electrolyte | Polymer | |
Matching pin | - | |
Amplifier | - | |
Temperature | -5 to 80°C - HT | |
Max Pressure | 6 bar (87psi) | |
Connector | BNC | |
Cable | 5m |