Hanna Instruments

Scientific Equipment

HI-9829-13042 Advanced Portable Multiparameter for pH, EC/Turbidity, DO

HI-9829-13042 Advanced Portable Multiparameter for pH, EC/Turbidity, DO

Hanna HI-9829-13042 Advanced Portable Multiparameter for pH, EC/Turbidity, DO

HI-9829-13042 Advanced Portable Multiparameter for pH, EC/Turbidity, DO

Fully integrated GPS system
2 year manufacturer warranty
£4,365.00 (exc. VAT) £5,238.00 (inc. VAT)

Designed for measuring water quality, this advanced professional multi-parameter measures up to 15 parameters and comes with an impressive range of features.

HI-9829-13042 Advanced Portable Multiparameter for pH, EC/Turbidity, DO information

Hanna's extended description

Features of the HI-9829-13042 include a fully integrated GPS system and an autonomous sensor for measuring pH, EC/Turbidity, DO and temperature on a 4m cable.

Main benefits

  • Display 1 - 12 parameters simultaneously from up to 15 selectable parameters
  • Fast Track data traceability and/or GPS positioning
  • Robust, intelligent logging probe with stainless steel tip and narrow diameter for wells and pipes
  • Field replaceable sensors
  • Clear screen with backlit graphics making it easy to read - even in low light levels
  • Rugged water resistant casing - ideal for harsh outdoor environments
  • Ergonomic design that is comfortable to hold and use
  • Context sensitive HELP menu
  • Information and instructions in plain text
  • Flexible user interface allows users to configure the meter to suit individual needs

Integrated GPS Positioning

Measurement locations can be tracked and logged with detailed co-ordinates to make it easier than ever to return to the same location for repeated data collection.

The GPS co-ordinates are displayed on screen with up to 10 measurement parameters and are recorded with logged data. Users can connect to GPS tracking software such as Google Maps to view locations where samples have been taken. Measurement information is then shown on the map on screen.

The GPS can be used in conjunction with Fast Track for ultimate tracking and data capture.


Fast Track Tag Identification System

Using an iButton data management system that connects to iButton tags installed at all sampling sites, users can log and label measurements with an alpha-numeric location ID.

Data is automatically captured by the meter and can be transferred accurately and efficiently to a PC for evaluation and analysis.


Intelligent Potentiometric Probe

Designed as a professional tool, the digital probe provides stable, noise-free sensor signal management without the need for pre-amplified pH sensors.


Key features include:

  • Conductivity/Turbidity sensor according to EN ISO 7027
  • Independent probe with data acquisition allowing on-site measurement, collection and data storage
  • Rugged and water resistant to IP67 with weighted end
  • Automatic recognition of sensors and electrodes
  • Field replaceable sensors

Applications

The HI-9829 is a high specification meter designed for taking professional field measurements in lakes, rivers and seas.


  • Aquaculture
  • Fish farming
  • Oceanography
  • Harbours and Ports
  • Surface, drainage and waste water testing


IP67 – Meter is dust-tight and protected against the effects of temporary immersion in water up to 1m.



PC Interface - Download test results to a computer programmed with HANNA’s Windows® compatible software (HI-92000), directly from equipment via a USB port or a RS232 cable outlet.

Specifications

Specifications list

Specifications

HI-9829 specification with/without GPS
Temperature Compensation Automatic from -5 to 55°C
Models with GPS 12 Channel Receiver, 10m range
Logging memory from meter 44,000 records
Logging interval 1 second to 3 hours
Computer interface USB (with HI-929829)
FastTracker™ TAG ID Yes
Waterproof protection IP67
Environment 0 to 50°C, RH 100%
Power supply 1.5 V Alkaline C cells (4)/1.2V NiMH rechargeable C Cells (4) USB, 12V power adapter
Dimensions 221 x 115 x 55mm
Weight 750g
Maximum Probe Immersion Depth – 20 metres

HI-9829 parameter specifications
pH Range 0.00 to 14.00 pH
pH Resolution 0.01 pH
pH Accuracy ±0.02 pH
pH Calibration automatic one, two, or three points with automatic recognition of five standard buffers (pH 4.01, 6.86, 7.01, 9.18, 10.01) or one custom buffer
mV Range ±600.0 mV
mV Resolution 0.1 mV
mV accuracy ±0.5 mV
ORP Range ±2000 mV
ORP Resolution 0.1 mV
ORP Accuracy ±1.0 mV
ORP Calibration automatic at one custom point (relative mV)
EC Range 0 to 200 mS/cm (absolute EC up to 400 mS/cm)
EC Resolution Manual: 1 µS/cm; 0.001 mS/cm; 0.01 mS/cm; 0.1 mS/cm; 1 mS/cm, Automatic: 1 µS/cm from 0 to 9999 µS/cm; 0.01 mS/cm from 10.00 to 99.99 mS/cm; 0.1 mS/cm from 100.0 to 400.0 mS/cm, Automatic (mS/cm): 0.001 mS/cm from 0.000 to 9.999 mS/cm; 0.01 mS/cm from 10.00 to 99.99 mS/cm; 0.1 mS/cm from 100.0 to 400.0 mS/cm
EC Accuracy ±1% of reading or ±1 µS/cm whichever is greater
EC Calibration automatic single point, with six standard solutions (84 µS/cm, 1413 µS/cm, 5.00 mS/cm, 12.88 mS/cm, 80.0 mS/cm, 111.8 mS/cm) or custom point
TDS Range 0 to 400000 ppm (mg/L); (the maximum value depends on the TDS factor)
TDS Resolution Manual: 1 ppm (mg/L); 0.001 ppt (g/L); 0.01 ppt (g/L); 0.1 ppt (g/L); 1 ppt (g/L), Automatic: 1 ppm (mg/L) from 0 to 9999 ppm (mg/L); 0.01 ppt (g/L) from 10.00 to 99.99 ppt (g/L); 0.1 ppt (g/L) from 100.0 to 400.0 ppt (g/L), Automatic ppt (g/L): 0.001 ppt (g/L) from 0.000 to 9.999 ppt (g/L); 0.01 ppt (g/L) from 10.00 to 99.99 ppt (g/L); 0.1 ppt (g/L) from 100.0 to 400.0 ppt (g/L)
TDS Accuracy ±1% of reading or ±1 ppm (mg/L), whichever is greater
TDS Calibration based on conductivity or salinity calibration
Resistivity Range 0 to 999999 Ω•cm; 0 to 1000.0 kΩ•cm; 0 to 1.0000 MΩ•cm
Resistivity Resolution dependent on resistivity reading
Resistivity Calibration based on conductivity calibration
Salinity Range 0.00 to 70.00 PSU
Salinity Resolution 0.01 PSU
Salinity Accuracy ±2% of reading or ±0.01 PSU whichever is greater
Salinity Calibration based on conductivity calibration
Seawater σ Range 0.0 to 50.0 σt, σ0, σ15
Seawater σ Resolution 0.1 σt, σ0, σ15
Seawater σ Accuracy ±1 σt, σ0, σ15
Seawater σ Calibration based on conductivity or salinity calibration
Ammonium-Nitrogen Range 0.02 to 200.0 ppm Am (as NH4 +-N)
Ammonium-Nitrogen Resolution 0.01 ppm to 1 ppm 0.1 ppm to 200.0 ppm
Ammonium-Nitrogen Accuracy ±5 % reading or 2 ppm
Ammonium-Nitrogen Calibration 1 or 2 point, 10 ppm and 100 ppm
Chloride Range 0.6 to 200.0 ppm Cl (as Cl- )
Chloride Resolution 0.01 ppm to 1 ppm 0.1 ppm to 200.0 ppm
Chloride Accuracy ±5 % of reading or 2 ppm
Chloride Calibration 1 or 2 point, 10 ppm and 100 ppm
Nitrate-Nitrogen Range 0.62 to 200.0 ppm Ni (as NO3 - -N)
Nitrate-Nitrogen Resolution 0.01 ppm to 1 ppm 0.1 ppm to 200 ppm
Nitrate-Nitrogen Accuracy ±5 % of reading or 2 ppm
Nitrate-Nitrogen Calibration 1 or 2 point, 10 ppm and 100 ppm
Turbidity Range 0.0 to 99.9 FNU; 100 to 1000 FNU
Turbidity Resolution 0.1 FNU from 0.0 to 99.9 FNU 1 FNU from 100 to 1000 FNU
Turbidity Accuracy ±0.3 FNU or ±2 % of reading, whichever is greater
Turbidity Calibration Automatic 1, 2, or, 3 points at 0, 20 and 200 FNU, or custom
DO Range 0.0 to 500.0%; 0.00 to 50.00 ppm (mg/L)
DO Resolution 0.1%; 0.01 ppm (mg/L)
DO Accuracy 0.0 to 300.0% saturation: ±1.5% of reading or ±1.0% saturation whichever is greater, 0.00 to 30.00 ppm (mg/L): ±1.5% of reading or ±0.10 ppm (mg/L), whichever is greater, 30.00 ppm (mg/L) to 50.00 ppm (mg/L): ±3% of reading, 300.0 to 500.0% saturation: ±3% of reading
DO Calibration automatic one or two points at 0 and 100% or one custom point
Atmospheric Pressure Range 450 to 850 mm Hg; 17.72 to 33.46 in Hg; 600.0 to 1133.2 mbar; 8.702 to 16.436 psi; 0.5921 to 1.1184 atm; 60.00 to 113.32 kPa
Atmospheric Pressure Resolution 0.1 mm Hg; 0.01 in Hg; 0.1 mbar; 0.001 psi; 0.0001 atm; 0.01 kPa
Atmospheric Pressure Accuracy ±3 mm Hg within ±15°C from the temperature during calibration
Atmospheric Pressure Calibration automatic at one custom point
Ordering Information The HI-9829 is supplied with a HI-710140 hard carrying case, HI-7612VDC adapter, HI-7698292 probe maintenance kit, HI-929829 PC application software, HI-7698291 USB cable (PC to meter), HI-710045 power supply cable, HI-710046 cigarette lighter power cable, HI-9829-25 quick calibration solution, HI-920005 iButton with holder (5), HI-7609829-1 pH/ORP sensor, HI-7609829-2 galvanic DO sensor, HI-7609829-3 EC sensor, HI-7698290 short calibration beaker, HI-7698295 short protective weighted shield, NiMH rechargeable “C” cell batteries (4), 12VDC power adapter, and instruction manual.
Probes selected with turbidity option will include the HI-7609829-4 EC/turbidity sensor, HI-7698293 long calibration beaker and HI-7698296 long protective weighted shield in place of the EC sensor, short calibration beaker and short protective shield. Probes selected with turbidity option will also include turbidity calibration solutions (230 ml); HI-9829-16 0 FNU, HI-769829-17 20 FNU, and HI-9829-18 200 FNU.

Spare Solutions
HI-9829-10 25 sachets 10ppm ammonia-nitrogen calibration solution
HI-9829-10/11 10 sachets each of 10ppm and 100ppm ammonia-nitrogen calibration solution
HI-9829-11 25 sachets 100ppm ammonia-nitrogen calibration solution
HI-9829-12 25 sachets 10ppm chloride calibration solution
HI-9829-12/13 10 sachets each of 10ppm and 100ppm chloride calibration solution
HI-9829-13 25 sachets 100ppm chloride calibration solution
HI-9829-14 25 sachets 10ppm nitrate-nitrogen calibration solution
HI-9829-14/15 10 sachets each of 10ppm and 100pm nitrate-nitrogen calibration solution
HI-9829-15 25 sachets 100ppm nitrate-nitrogen calibration solution

Downloadable Manuals

Manuals

Download PDF

Product Literature

literature - PDF

Hanna Instruments Literature

Questions & Answers

Got a question to ask? Use this form to submit your question.
Question will be displayed after moderation.
Name* Product Question*
No questions yet?
Don 't be shy, be the first to ask.