Hanna Instruments

Scientific Equipment

Safety Data Sheets

SDS search & quick download

Free Delivery

Most of mainland GB for orders over £50. Other charges may apply - see more

Hanna Care

We take the long view, looking after existing customers as much as delighting new ones!

Certificates of Analysis

Download a copy of the certificate of analysis for your solution

HI-5321 Research Grade EC/TDS Meter with USP

HI-5321 Research Grade EC/TDS Meter with USP

Hanna HI-5321 Research Grade EC/TDS Meter with USP

HI-5321 Research Grade EC/TDS Meter with USP

Complete with EC probe
2 year manufacturer warranty
£882.80 (exc. VAT) £1,059.36 (inc. VAT)

HI-5321 can be used to perform all 3 stages of USP <645> method required for conductivity measurement of pure and ultra-pure water.

PC Interface - Download test results to a computer programmed with HANNA’s Windows® compatible software (HI-92000), directly from equipment via a USB port or a RS232 cable outlet.

Replaceable Electrode – No need to replace the whole meter when the electrode needs replacing. Spare and replacement electrodes are available for this product.

Frequently Bought Together

Price for all:
£908.80

HI-5321 Research Grade EC/TDS Meter with USP information

Hanna's extended description

The instrument provides clear directions on how to perform each testing stage and automatically monitors the temperature, conductivity and stability during testing and determines whether a sample is within USP limits.

This instrument is equipped with auto standard recognition and can support custom calibration solutions. Up to a four point calibration can be obtained for enhanced accuracy over an extended measuring range.

The enhanced warning system is in place to alert users when measuring outside the calibration range. This system will also remind the user when a new calibration is due.

Features also include real-time graphic displays and on-screen GLP data. On-board contextual help can be accessed from any mode via the HELP button.

Up to 10 profiles can be saved and recalled eliminating the need to reconfigure each time when a different electrode is used.

Three selectable logging modes are available: Automatic, Manual and AutoHold logging. Up to 100 logging lots with 10,000 records per lot can be stored for automatic or manual modes along with up to 200 USP reports.

All features are available in English, Italian, Portuguese and Spanish. Connect to a PC via the USB or RS232 ports with the use of HI-92000 Windows® compatible software.

Ideal for

  • pharmaceutical
  • advanced research
  • vaccine development
  • environmental laboratories

Supplied with

  • HI-76312 conductivity probe
  • HI-76404N electrode holder
  • 12VDC power adapter
  • Instruction manual

Specifications

Specifications list

Specifications

EC - Range 0.000 to 9.999 µS/cm; 10.00 to 99.99 µS/cm; 100.0 to 999.9 µS/cm; 1.000 to 9.999 mS/cm; 10.00 to 99.99 mS/cm; 100 0 to 1000.0 mS/cm absolute EC*
 Resolution 0.001µS/cm; 0.01µS/cm; 0.1µS/cm; 0.001mS/cm; 0.01mS/cm; 0.1mS/cm
 Accuracy ±1% of reading (±0.01µS/cm)
 Cell Constant 0.0500 to 200.00
 Cell Type 4 cells
 Calibration automatic standard recognition, user standard single point/multi-point calibration
 Calibration Reminder yes
  Temperature Coefficient 0.00 to 10.00%/°C
  Temperature Compensation disabled, linear and non-linear (natural water)
 Reference Temperature 5.0 to 30.0°C
 Profiles up to 10, 5 each channel
 USP Compliant yes
TDS - Range 0.000 to 9.999 ppm; 10.00 to 99.99 ppm; 100.0 999.9 ppm; 1.000 to 9.999 ppt; 10.00 to 99.99 ppt; 100.0 to 400.0 ppt actual TDS (with 1.00 factor)
 Resolution 0.001 ppm; 0.01 ppm; 0.1 ppm; 0.001 ppt; 0.01 ppt; 0.1 ppt; 0.001 ppm; 0.001 ppm; 0.01 ppm; 0.1 ppm; 0.001 ppt; 0.01 ppt; 0.1 ppt
 Accuracy ±1% of reading (±0.01 ppm)
Resistivity - Range 1.0 to 99.9 Ω • cm; 100 to 999 Ω •cm; 1.00 to 9.99 K Ω •cm; 10.0 to 99.9 K Ω • cm; 100 to 999 K Ω • cm; 1.00 to 9.99 M Ω •cm; 10.0 to 100.0 M Ω • cm
 Resolution 0.1 Ω • cm; 1 Ω • cm; 0.01 K Ω • cm; 0.1 K Ω • cm; 1 K Ω • cm; 0.01 M Ω • cm; 0.1 M Ω • cm
 Accuracy ±2% of reading (± 1 Ω • cm)
Salinity - Range practical scale: 0.00 to 42.00 psu; natural sea water scale: 0.00 to 80.00 ppt; percent scale: 0.0 to 400.0% NaCl
  Resolution 0.01 for practical scale/natural sea water scale; 0.1% for percent scale
 Accuracy ±1% of reading
  Calibration percent scale-one point (with HI-7037 standard)
Temperature - Range -20.0 to 120°C
 Resolution 0.1°C
 Accuracy ±0.2°C (without probe)
Temperature Compensation Linear and non-linear (natural water)
EC Probe HI-76312 platinum, four ring EC/TDS probe with 1m cable (included)
GLP cell constant, reference temperature/coefficient, calibration points, calibration time stamp, probe offset for conductivity
Logging record: 100,000 data point storage/channel, up to 100 lots with max, 50,000 records-lot; Interval: settable between 1 second and max log time of 180 minutes;
Display colour graphic LCD with on-screen help, graphing, and custom colour configuration
PC Connection USB
Power Supply 12VDC adapter (included)
Environment 0 to 50°C; RH max 95% non-condensing
Dimensions/Weight 160 x 231 x 94 mm / 1.2kg

Questions & Answers

Got a question to ask? Use this form to submit your question.
Question will be displayed after moderation.
Name* Product Question*
No questions yet?
Don 't be shy, be the first to ask.