HI-2221 Calibration Check pH Bench Meter information
Hanna's extended description
The HI-2221-02 is a pH bench top meter features our exclusive Calibration Check diagnostics for both pH electrodes and buffer solutions during calibration. This instrument compares the characteristics of the pH electrode from one calibration to the next. In the case of large variances in the electrode condition, it alerts the user that the electrode needs to be properly cleaned prior to calibration and measuring.
The second feature is to detect if the calibration buffer solution is contaminated. After calibration, the probe condition is evaluated and an indicator is displayed informing the user of the overall pH electrode status.
This instrument can measure using ORP electrodes (pH channel input), due to its capability to measure mV with a resolution up to 0.1 mV.
A properly manufactured and maintained pH electrode will retain its measuring integrity for a long time. As a result of many factors such as age, poor maintenance or improper handling, any electrode in time will lose its integrity. The most common cause for pH measurement inaccuracies is an unclean or improperly cleaned electrode. This is very important to note, because during calibration, the instrument assumes that the electrode is clean and that the standardisation curve created during the calibration process will remain a valid reference until the next calibration.
Key features

Replaceable Electrode – No need to replace the whole meter when the electrode needs replacing. Spare and replacement electrodes are available for this product.

PC Interface (optional) - Download test results to a computer programmed with HANNA’s Windows® compatible software (HI-92000), directly from equipment via a USB port or a RS232 cable outlet.
Specifications
Specifications list
Specifications
Range pH | -2.00 to +16.00 pH | |
mV | ± 699.9 mV; ± 2000 mV | |
Temperature | -20.0 to 120.0 °C | |
Resolution pH | 0.01 pH | |
mV | 0.1 mV (± 699.9 mV); 1 mV (± 2000 mV) | |
Temperature | 0.1 °C | |
Accuracy pH | ± 0.01 pH | |
mV | ± 0.2 mV (± 699.9 mV); ± 1 mV (± 2000 mV) | |
Temperature | ± 0.2 °C (excluding probe error) | |
pH Calibration | Up to 5 point calibration, 7 standard buffers available (1.68, 4.01, 6.86, 7.01, 9.18, 10.01, 12.45) | |
Temperature Compensation | Manual or automatic from -20 to 120°C | |
Temperature Probe | HI 1131P glass body, single junction, refillable with BNC connection and 1 m (3.3') Cable HI 7662 stainless steel probe | |
PC Connection | USB | |
Data Logging | 100 points | |
Dimensions | 235 x 222 x 109 mm (9.2 x 8.7 x 4.3") | |
Weight | 1.3 kg (2.9 lb.) |
Downloadable Manuals
Manuals
MAN_2221_2223_06_19.pdfQuestions & Answers
Question will be displayed after moderation.