HI-96729 Fluoride Portable Photometer

HI-96729 Fluoride Portable Photometer

Hanna HI-96729 Fluoride Portable Photometer

HI-96729 Fluoride Portable Photometer

Complete with cuvettes
2 year manufacturer warranty
£420.50 (exc. VAT) £504.60 (inc. VAT)
The Hanna HI-96729 is an auto-diagnostic portable photometer for accurately measuring fluoride from 0.00 to 2.00 mg/L.
Water Resistant – Can withstand temporary exposure to water, but should not be submerged or washed under running water.
Battery powered – Allows use away from a power source. Batteries are included with all battery operated products.

Frequently Bought Together

Price for all:
£452.00

HI-96729 Fluoride Portable Photometer information

Hanna's extended description

Fluoride, Low Range Portable Photometer

The HI-96729 meter measures the fluoride content in the 0.00 to 2.00 mg/L (ppm) range, in drinking, surface and waste waters. The amount of reagent is precisely dosed by use of the supplied automatic pipette for maximum repeatability.

This meter uses an exclusive positive-locking system to ensure that the cuvette is in the same position every time it is placed into the measurement cell. It is designed to fit cuvettes with a larger neck making it easier to add both sample and reagents. The cuvettes are made from special optical glass to obtain best results.

Also available as a full kit HI-96729C.


Ideal for

  • Water: drinking, surface and waste water

Key features

  • Clear display messages make the instrument quick and easy to operate
  • The powerful Cal Check calibration function means that there is no need to return the instrument to the laboratory to check its efficiency. Samples can be checked and validation confirmed at the press of a button
  • Positive locking system ensures the cuvette is located in the same position each and every time is it placed in the measurement cell
  • Auto shut off feature turns meter off after 10 minutes of non-use to save battery life
  • Fully water resistant
  • Clear easy to use function buttons
  • Direct readout of parameter measured

Designed for precision

With the Hanna calibration check validation function, users can verify the performance of the meter quickly and easily at any time. Using Hanna NIST traceable standards, it takes just a few moments to check the accuracy of the instrument and to recalibrate as and when necessary.

All new instruments are factory calibrated at source and the electronic and optical design of the Hanna photometers minimises the need for frequent calibration. The Hanna calibration check further minimises the frequency with which instruments need to be returned to the calibration laboratory for full servicing.


Order information

  • Meter only (HI -96729): 2 glass cuvettes, 9V battery

NB: reagents ordered separately



Specifications

Specifications list

Specifications

Parameter Fluoride, Low Range
Range 0.00 to 2.00 mg/L
Accuracy & resolution ±0.03 mg/L   1.00 mg/L
Filter Narrow band @ 575 nm
Method Adaptation of the EPA method 340.1 and SPADNS method.
Environment 0 to 50°C (32 to 122°F); Max 95% RH non-condensing
Battery Type / Life (1) 9V
Auto Shut-off After 10 min. of non-use in measurement mode
Dimensions 193 x 104 x 69 mm
Weight 360g
Ordering information HI-96729 comprises photometer, sample cuvettes (2) with caps, 9V battery, instrument quality certificate and instruction manual
Reagents and standards (sold separately) HI-93703-53 reagent for reducing chlorine concentration; HI-96729-11 CAL CheckTM standard cuvettes; HI-93729-01 reagents for 100 tests; HI-93729-03 regents for 300 meters
PLEASE NOTE: In some cases, samples may contain other parameters which may interfere with results
Interference may be caused by: Chlorine interferes at all levels: to eliminate Cl2 interference add to the 10 mL unreacted sample, 1 drop of the optional HI-93703-53 Chlorine Remover, for each 2 mg/L of chlorine.
• Positive error: Chloride above 7000 mg/L; Phosphate, ortho above 16 mg/L; Sodium hexametaphosphate above 1.0 mg/L; Sulphate above 200 mg/L
• Negative error: Alkalinity (as CaCO3) above 5000 mg/L; Aluminium above 0.1 mg/L; Iron, ferric above 10 mg/L

Downloadable Manuals

Manuals

Download PDF

Product Literature

literature - PDF

Hanna Instruments Literature

Questions & Answers

Got a question to ask? Use this form to submit your question.
Question will be displayed after moderation.
Name* Product Question*
No questions yet?
Don 't be shy, be the first to ask.