Hanna HI-5321 Research Grade EC/TDS Meter with USP

HI-5321-02

HI-5321 can be used to perform all 3 stages of USP method required for conductivity measurement of pure and ultra-pure water.

£1,048.80 (exc.VAT)
£1,258.56 (tax incl.)
Order now for 2 weeks delivery
Accessories
Reference
Price
Quantity
Full Description Specification Manuals

HI-5321 can be used to perform all 3 stages of USP <645> method required for conductivity measurement of pure and ultra-pure water. The instrument provides clear directions on how to perform each testing stage and automatically monitors the temperature, conductivity and stability during testing and determines whether a sample is within USP limits.

This instrument is equipped with auto standard recognition and can support custom calibration solutions. Up to a four point calibration can be obtained for enhanced accuracy over an extended measuring range.

The enhanced warning system is in place to alert users when measuring outside the calibration range. This system will also remind the user when a new calibration is due.

Features also include real-time graphic displays and on-screen GLP data. On-board contextual help can be accessed from any mode via the HELP button.

Up to 10 profiles can be saved and recalled eliminating the need to reconfigure each time when a different electrode is used.

Three selectable logging modes are available: Automatic, Manual and AutoHold logging. Up to 100 logging lots with 10,000 records per lot can be stored for automatic or manual modes along with up to 200 USP reports.

All features are available in English, Italian, Portuguese and Spanish. Connect to a PC via the USB or RS232 ports with the use of HI-92000 Windows® compatible software.

Ideal for

  • pharmaceutical advanced research
  • vaccine development
  • environmental laboratories

Supplied with:

  • HI-76312 conductivity probe
  • HI-76404N electrode holder
  • 12VDC power adapter
  • Instruction manual

Read More

EC - Range0.000 to 9.999 µS/cm; 10.00 to 99.99 µS/cm; 100.0 to 999.9 µS/cm; 1.000 to 9.999 mS/cm; 10.00 to 99.99 mS/cm; 100 0 to 1000.0 mS/cm absolute EC*
 Resolution0.001µS/cm; 0.01µS/cm; 0.1µS/cm; 0.001mS/cm; 0.01mS/cm; 0.1mS/cm
 Accuracy±1% of reading (±0.01µS/cm)
 Cell Constant0.0500 to 200.00
 Cell Type4 cells
 Calibrationautomatic standard recognition, user standard single point/multi-point calibration
 Calibration Reminderyes
  Temperature Coefficient0.00 to 10.00%/°C
  Temperature Compensationdisabled, linear and non-linear (natural water)
 Reference Temperature5.0 to 30.0°C
 Profilesup to 10, 5 each channel
 USP Compliantyes
TDS - Range0.000 to 9.999 ppm; 10.00 to 99.99 ppm; 100.0 999.9 ppm; 1.000 to 9.999 ppt; 10.00 to 99.99 ppt; 100.0 to 400.0 ppt actual TDS (with 1.00 factor)
 Resolution0.001 ppm; 0.01 ppm; 0.1 ppm; 0.001 ppt; 0.01 ppt; 0.1 ppt; 0.001 ppm; 0.001 ppm; 0.01 ppm; 0.1 ppm; 0.001 ppt; 0.01 ppt; 0.1 ppt
 Accuracy±1% of reading (±0.01 ppm)
Resistivity - Range1.0 to 99.9 Ω • cm; 100 to 999 Ω •cm; 1.00 to 9.99 K Ω •cm; 10.0 to 99.9 K Ω • cm; 100 to 999 K Ω • cm; 1.00 to 9.99 M Ω •cm; 10.0 to 100.0 M Ω • cm
 Resolution0.1 Ω • cm; 1 Ω • cm; 0.01 K Ω • cm; 0.1 K Ω • cm; 1 K Ω • cm; 0.01 M Ω • cm; 0.1 M Ω • cm
 Accuracy±2% of reading (± 1 Ω • cm)
Salinity - Rangepractical scale: 0.00 to 42.00 psu; natural sea water scale: 0.00 to 80.00 ppt; percent scale: 0.0 to 400.0% NaCl
  Resolution0.01 for practical scale/natural sea water scale; 0.1% for percent scale
 Accuracy±1% of reading
  Calibrationpercent scale-one point (with HI-7037 standard)
Temperature - Range-20.0 to 120°C
 Resolution0.1°C
 Accuracy±0.2°C (without probe)
Temperature CompensationLinear and non-linear (natural water)
EC ProbeHI-76312 platinum, four ring EC/TDS probe with 1m cable (included)
GLPcell constant, reference temperature/coefficient, calibration points, calibration time stamp, probe offset for conductivity
Loggingrecord: 100,000 data point storage/channel, up to 100 lots with max, 50,000 records-lot; Interval: settable between 1 second and max log time of 180 minutes;
Displaycolour graphic LCD with on-screen help, graphing, and custom colour configuration
PC ConnectionUSB
Power Supply12VDC adapter (included)
Environment0 to 50°C; RH max 95% non-condensing
Dimensions/Weight160 x 231 x 94 mm / 1.2kg