One meter can be used for both photometric and pH measurements. HI-83399 is a compact, multi-parameter photometer for measuring key water and wastewater quality parameters.

40 key water quality parameters
pH
COD

£1,841.40 (exc.VAT)
£2,209.68 (tax incl.)
Full Description Specification

HI-83399 Water and waste water multi-parameter with COD photometer and pH meter

Advanced optical system - Unparalleled performance from a benchtop photometer.

One meter can be used for both photometric and pH measurements. HI-83399 is a compact, multi-parameter photometer for measuring key water and wastewater quality parameters. The meter is one of the most advanced photometers available, with an innovative optical design that utilises a reference detector and focusing lens to eliminate errors from changes in the light source and imperfections in the glass cuvette.

40 key water and wastewater quality parameters with 73 different methods covering multiple ranges are programmed into the meter. Wastewater treatment digestion parameters include COD, Total Nitrogen and Total Phosphorous, which are important for monitoring nutrient removal. The HI-83399 also offers an absorbance measurement mode for performance verification and for users who would like to develop their own concentration versus absorbance curves.

The Chemical Oxygen Demand (COD) parameter is included for industrial and municipal wastewater treatment. The Phosphorous and Nitrogen parameters included are beneficial to municipal wastewater treatment customers that need to monitor their biological and chemical nutrient removal process. This photometer features an innovative optical system that uses LEDS, narrow band interference filters, focusing lens and both a silicon photodetector for absorbance measurement and a reference detector to maintain a consistent light source ensures accurate and repeatable photometric readings every time.

A digital pH electrode input is provided allowing the user to measure pH by a traditional glass electrode.

Two USB ports are provided for transferring data to a flash drive or computer and to use as a power source for the meter. For added convenience and portability the meter can also operate on an internal 3.7 VDC Lithium-polymer rechargeable battery.

The HI-83399 offers an absorbance measuring mode that allows for CAL Check standards to be used to validate the performance of the system. The absorbance mode allows the user to select one of the 5 wavelengths of light (420 nm, 466 nm, 525 nm, 575 nm, and 610 nm) to measure and plot their own concentration versus absorbance mode. This is useful for users with their own chemical method and for education to teach the concept of absorbance by using the Beer-Lambert Law.

Key features:

  • Advanced optical system with brighter, long-lasting LED light source
  • Backlit 128 x 64 Pixel Graphic Liquid Crystal Display
  • Built-in Reaction Timer for Photometric Measurements
  • Absorbance mode - concentration versus absorbance
  • Units of Measurement plus chemical form displayed
  • Result conversion at the touch of a button
  • Cuvette cover
  • pH and temperature measurement with a single probe
  • Good Laboratory Practice (GLP) - track calibration information including date, time, buffers used, offset and slope for traceability
  • CAL Check alerts user to potential problems during the calibration process
  • Data Logging - Up to 1000 photometric and pH readings can be stored
  • Logged readings can be quickly and easily transferred to a flash drive or to a PC. Data is exported as a .CSV file for use with spreadsheet programs
  • Battery Status Indicator
  • Error Messages

High Efficiency LED Light Source

An LED light source offers superior performance compared to a tungsten lamp. LEDs have a much higher luminous efficiency, providing more light while using less power. They also produce very little heat, which could otherwise affect the optical components and electronic stability. LEDs are available in a wide array of wavelengths, whereas tungsten lamps are supposed to be white light (all wavelengths of visible light) but actually have a poor blue/violet light output.

High Quality Narrow Band Interference Filters

The narrow band interference filter not only ensure greater wavelength accuracy (+/- 1 nm) but are extremely efficient. The filters used allow up to 95% of the light from the LED to be transmitted as compared to other filters that are only 75% efficient. The higher efficiency allows for a brighter, stronger light source. The end result is higher measurement stability and less wavelength error.

Reference Detector for a Stable Light Source

A beam splitter is used as part of the internal reference system of the HI-83300 photometer. The reference detector compensates for any drift due to power fluctuations or ambient temperature changes. Now you can rely on a stable source of light between your blank (zero) measurement and sample measurement.

Large Cuvette Size

The sample cell of the HI-83308 fits a round, glass cuvette with a 25 mm path length. Along with the advanced optical components, the larger size of the cuvette greatly reduces errors in rotation from the indexing mark of the cuvettes. The relatively long path length of the sample cuvette allows the light to pass through more of the sample solution, ensuring accurate measurements even in low absorbance samples.

Focusing Lens for Greater Light Yield

Adding a focusing lens to the optical path allows for the collection of all of the light that exits the cuvette and focusing the light on the silicon photo detector. This novel approach to photometric measurements cancels the errors from imperfections and scratches present in the glass cuvette eliminating the need to index the cuvette.

Order COD reagents here

Read More

Input Channels1 pH electrode input and 5 photometer wavelengths
pH ElectrodeDigital pH electrode (not included)
Logging typeLog on demand with user name and sample ID optional input
Logging Memory1000 readings
ConnectivityUSB-A host for flash drive; micro-USB-B for power and computer connectivity
GLPCalibration data for connected pH electrode
Display128 x 64 pixel LCD with backlight
Battery Type/Life3.7 VDC Li-polymer rechargeable battery/ >500 photometric measurements or 50 hours of continuous pH measurement
Power supply5 VDC USB 2.0 power adapter with USB-A to micro-USB-B cable (included)
Environment0 to 50.0°C (32 to 122.0°F; 0 to 95% RH, non-condensing
Dimensions206 x 177 x 97mm (8.1 x 7.0 x 3.8")
Weight1.0 kg (2.2lbs)
Photometer/Colorimeter Light Source5 LEDs with 420nm, 466nm, 525nm, 575nm, and 610nm narrow band interference filters
Photometre/Colorimeter Light DetectorSilicon Photodetector
Bandpass filter Bandwidth8nm
Bandpass filter wavelength accruacy+-1nm
Cuvette TypeRound, 24.6mm
Number of Methods128 Max
Ordering InformationHI-83399 is supplied with sample cuvettes and caps (x4), cloth for wiping cuvettes, USB to micro USB cable connector, power adapter and instruction manual.
ALKALINITY
Alkalinity RangeFreshwater: 0 to 500 mg/L (as CaCO3) Marine: 0 to 300 mg/L (as CaCO3)
Alkalinity Resolution1 mg/L
Alkalinity Accuracy±5 mg/L ±5% of reading
Alkalinity MethodBromocresol green method
ALUMINIUM
Aluminium Range0.00 to 1.00 mg/L (as Al3+)
Aluminium Resolution0.01 mg/L
Aluminium Accuracy±0.04 mg/L ±4% of reading
Aluminium MethodAdaptation of the aluminon method
AMMONIA
Ammonia RangeLow Range: 0.00 to 3.00 mg/L (as NH3-N) Medium Range: 0.00 to 10.00 mg/L (as NH3-N) High Range: 0.0 to 100.0 mg/L (as NH3-N)
Ammonia Resolution0.01 mg/L; 0.1 mg/L
Ammonia AccuracyLow Range: ±0.04 mg/L ±4% of reading Medium Range: ±0.05 mg/L ±5% of reading High range: ±0.5 mg/L ±5% of reading at 25 °C
Ammonia MethodAdaptation of the ASTM Manual of Water and Environmental Technology, D1426-92, Nessler method
BROMINE
Bromine Range0.00 to 8.00 mg/L (as Br2)
Bromine Resolution0.01 mg/L
Bromine Accuracy±0.08 mg/L ±3% of reading
Bromine MethodAdaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, DPD method.
CALCIUM
Calcium RangeFreshwater: 0 to 400 mg/L (as Ca2+) Marine: 200 to 600 mg/L (as Ca2+)
Calcium Resolution1 mg/L
Calcium AccuracyFreshwater: ±10 mg/L ±5% of reading Marine: ±6% of reading
Calcium MethodFreshwater: Adaptation of the Oxalate method Seawater: Adaptation of the Zincon method
CHLORIDE
Chloride Range0.0 to 20.0 mg/L (as Cl⁻)
Chloride Resolution0.1 mg/L
Chloride Accuracy±0.5 mg/L ±6% of reading at 25 °C
CHLORINE DIOXIDE
Chlorine Dioxide Range0.00 to 2.00 mg/L (as ClO2)
Chlorine Dioxide Resolution0.01 mg/L
Chlorine Dioxide Accuracy±0.10 mg/L ±5% of reading
Chlorine Dioxide MethodAdaptation of the Chlorophenol Red method.
CHLORINE
Free Chlorine Range0.00 to 5.00 mg/L (as Cl2) Ultra Low Range: 0.000 to 0.500 mg/L (as Cl2)
Free Chlorine Resolution0.01 mg/L Ultra Low Range: 0.001 mg/L
Free Chlorine Accuracy±0.03 mg/L ±3% of reading Ultra Low Range: ±0.020 mg/L ±3% of reading
Total Chlorine Range0.00 to 5.00 mg/L (as Cl2) Ultra Low Range: 0.000 to 0.500 mg/L (as Cl2) Ultra High Range: 0 to 500 mg/L (as Cl2)
Total Chlorine Resolution0.01 mg/L; 0.001 mg/L; 1 mg/L
Total Chlorine Accuracy±0.03 mg/L ±3% of reading Ultra Low Range: ±0.020 mg/L ±3% of reading Ultra High Range: ±3 mg/L ±3% of reading
Chlorine MethodAdaptation of the EPA 330.5 DPD method Free Chlorine (ULR) & Total Chlorine (UHR): Adaptation of the Standard Methods for Examination of Water and Wastewater, 20th edition, 4500-Cl
CHROMIUM, HEXAVALENT
Chromium, Hexavalent RangeLow Range: 0 to 300 μg/L (as Cr6+) High Range: 0 to 1000 μg/L (as Cr6+)
Chromium, Hexavalent Resolution1 μg/L
Chromium, Hexavalent AccuracyLow Range: ±1 μg/L ±4% of reading High Range: ±5 μg/L ±4% of reading
Chromium, Hexavalent MethodAdaptation of the ASTM Manual of Water and Environmental Technology, D1687-92, Diphenylcarbohydrazide method.
COD
COD Low, Medium & High RangeLow Range: 0 to 150 mg/L (as O2) Medium Range: 0 to 1500 mg/L (as O2) High Range: 0 to 15000 mg/L (as O2
Resolution1mg/L
AccuracyLow Range: ±5 mg/L ±4% of reading Medium Range: ±15 mg/L ±4% of reading High Range: ±150 mg/L or ±2% of reading
MethodLow and medium range: dichromate mercury-free High Range: dichromate
COLOUR, WATER
Colour, Water Range0 to 500 PCU (Platinum Cobalt Units)
Colour, Water Resolution1 PCU
Colour, Water Accuracy±10 PCU ±5% of reading
Colour, Water MethodAdaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, Colorimetric Platinum Cobalt method.
COPPER
Copper RangeLow Range: 0.000 to 1.500 mg/L (as Cu2+) High range: 0.00 to 5.00 mg/L (as Cu2+)
Copper Resolution0.001 mg/L; 0.01 mg/L
Copper AccuracyLow Range: ±0.01 mg/L ±5% of reading High Range ±0.02 mg/L ±4% of reading
Copper MethodAdaptation of the EPA bicinchoninate method
CYANURIC ACID
Cyanuric Acid Range0 to 80 mg/L (as CYA)
Cyanuric Acid Resolution1 mg/L
Cyanuric Acid Accuracy±1 mg/L ±15% of reading
Cyanuric Acid MethodAdaptation of the turbidimetric method
FLUORIDE
Fluoride RangeLow Range: 0.00 to 2.00 mg/L (as F-) High range: 0.0 to 20.0 mg/L (as F-)
Fluoride Resolution0.01 mg/L; 0.1 mg/L
Fluoride AccuracyLow Range: ±0.03 mg/L ±3% of reading High Range: ±0.5 mg/L ±3% of reading
Fluoride MethodAdaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, SPADNS method
HARDNESS, CALCIUM
Hardness, Calcium Range0.00 to 2.70 mg/L (as CaCO3)
Hardness, Calcium Resolution0.01 mg/L
Hardness, Calcium Accuracy±0.11 mg/L ±5% of reading
Hardness, Calcium MethodAdaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, Calmagite method
HARDNESS, MAGNESIUM
Hardness, Magnesium Range0.00 to 2.00 mg/L (as CaCO3)
Hardness, Magnesium Resolution0.01 mg/L
Hardness, Magnesium Accuracy±0.11 mg/L ±5% of reading
Hardness, Magnesium MethodAdaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, EDTA Colorimetric method (Calmagite)
HARDNESS, TOTAL
Hardness, Total RangeLow Range: 0 to 250 mg/L (as CaCO3) Medium Range: 200 to 500 mg/L (as CaCO3) High Range: 400 to 750 mg/L (as CaCO3)
Hardness, Total Resolution1 mg/L
Hardness, Total AccuracyLow Range: ±5 mg/L ±4% of reading Medium Range: ±7 mg/L ±3% of reading High Range: ±10 mg/L ±2% of reading
Hardness, Total MethodCalmagite
HYDRAZINE
Hydrazine Range0 to 400 μg/L (as N2H4)
Hydrazine Resolution1 μg/L
Hydrazine Accuracy±4% of full scale reading
Hydrazine MethodAdaptation of the ASTM Manual of Water and Environmental Technology, method D1385-88, p-Dimethylaminobenzaldehyde method
IODINE
Iodine Range0.0 to 12.5 mg/L (as I2)
Iodine Resolution0.1 mg/L
Iodine Accuracy±0.1 mg/L ±5% of reading
Iodine MethodAdaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, DPD method
IRON
Iron RangeLow Range: 0.000 to 1.600 mg/L (as Fe) High Range: 0.00 to 5.00 mg/L (as Fe)
Iron Resolution0.001 mg/L; 0.01 mg/L
Iron AccuracyLow Range: ±0.01 mg/L ±8% of reading High Range: ±0.04 mg/L ±2% of reading
Iron MethodLow Range: Adaptation of the TPTZ Method High Range: Adaptation of the EPA Phenanthroline method 315B, for natural and treated waters
MAGNESIUM
Magnesium Range0 to 150 mg/L (as Mg2+)
Magnesium Resolution1 mg/L
Magnesium Accuracy±5 mg/L ±3% of reading
Magnesium MethodAdaptation of the Calmagite method
MANGANESE
Manganese RangeLow Range: 0 to 300 µg/L (as Mn) High range: 0.0 to 20.0 mg/L (as Mn)
Manganese Resolution1 µg/L; 0.1 mg/L
Manganese AccuracyLow Range: ±10 µg/L ±3% of reading High Range: ±0.3 mg/L ±5% of reading
Manganese MethodLow Range: Adaptation of the PAN Method High Range: Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, Periodate method
MOLYBDENUM
Molybdenum Range0.0 to 40.0 mg/L (as Mo6+)
Molybdenum Resolution0.1 mg/L
Molybdenum Accuracy±0.3 mg/L ±5% of reading
Molybdenum MethodAdaptation of the mercaptoacetic acid method
NICKEL
Nickel RangeLow Range: 0.000 to 1.000 mg/L (as Ni) High Range: 0.00 to 7.00 g/L (as Ni)
Nickel Resolution0.001 mg/L; 0.01 g/L
Nickel AccuracyLow range: ±0.010 mg/L ±7% of reading High Range: ±0.07 g/L ±4% of reading
Nickel MethodLow Range: Adaptation of the PAN method High Range: Adaptation of the photometric method
NITRATE
Nitrate Range0.0 to 30.0 mg/L (as NO3-- N)
Nitrate Resolution0.1 mg/L
Nitrate Accuracy±0.5 mg/L ±10% of reading
Nitrate MethodAdaptation of the cadmium reduction method
NITRITE
Nitrite RangeFreshwater Low Range: 0 to 600 μg/L (as NO2--N) High Range: 0 to 150 mg/L (as NO2-) Marine Ultra Low Range: 0 to 200 μg/L (as NO2--N)
Nitrite ResolutionFreshwater: 1 μg/L; 1 mg/L Marine: 0.1 mg/L
Nitrite AccuracyFreshwater Low Range: ±20 μg/L ±4% of reading High Range: ±4 mg/L ±4% of reading Marine: ±0.5 mg/L ±10% of reading
Nitrite MethodAdaptation of the EPA Diazotization method 354.1; HR Ferrous Sulphate; Cadium Reduction
NITROGEN
Nitrogen, Total Low & High RangeTotal Low Range: 0.0 to 25.0 mg/L (as NO3--N) High Range: 0 to 150 mg/L (as N)
Resolution1mg/L
AccuracyLow Range: ±1.0 mg/L ±5% of reading High Range: ±3 mg/L or ±4% of reading
MethodChromotropic acid
OXYGEN, DISSOLVED
Oxygen, Dissolved Range0.0 to 10.0 mg/L (as O2)
Oxygen, Dissolved Resolution0.1 mg/L
Oxygen, Dissolved Accuracy±0.4 mg/L ±3% of reading
Oxygen, Dissolved MethodAdaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, Azide modified Winkler method
OXYGEN, SCAVENGER
Oxygen, Scavenger Range0 to 1000 μg/L (as DEHA) 0.00 to 1.50 mg/L (as Carbohydrazide) 0.00 to 2.50 mg/L (as Hydroquinone) 0.00 to 4.50 mg/L (as ISO-ascorbic acid)
Oxygen, Scavenger Resolution1 μg/L (DEHA); 0.01 mg/L
Oxygen, Scavenger Accuracy±5 μg/L ±5% of reading
Oxygen, Scavenger MethodAdaptation of the iron reduction method
OZONE
Ozone Range0.00 to 2.00 mg/L (as O3)
Ozone Resolution0.01 mg/L
Ozone Accuracy±0.02 mg/L ±3% of reading
Ozone MethodColorimetric DPD Method
pH
pH Range6.5 to 8.5 pH
pH Resolution0.1 pH
pH Accuracy±0.1 pH
pH Methodphenol red
PHOSPHATE
Phosphate RangeFreshwater Low Range: 0.00 to 2.50 mg/L (as PO43-) High range: 0.0 to 30.0 mg/L (as PO43-) Seawater Ultra Low Range: 0 to 200 μg/L (as P)
Phosphate ResolutionFreshwater: 0.01 mg/L; 0.1 mg/L Seawater: 1 μg/L
Phosphate AccuracyFreshwater Low Range: ±0.04 mg/L ±4% of reading High Range: ±1 mg/L ±4% of reading Seawater Ultra Low Range: ±5 μg/L ±5% of reading
Phosphate MethodFreshwater Low Range: Adaptation of the Ascorbic Acid method Freshwater High Range and Seawater Ultra Low Range: Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, Amino Acid method
POTASSIUM
Potassium Range0.0 to 20.0 mg/L (as K)
Potassium Resolution0.1 mg/L
Potassium Accuracy±3.0 mg/L ±7% of reading
Potassium MethodAdaptation of the Turbidimetric Tetraphenylborate method
SILICA
Silica RangeLow Range: 0.00 to 2.00 mg/L (as SiO2) High Range: 0 to 200 mg/L (as SiO2)
Silica Resolution0.01 mg/L; 1 mg/L
Silica AccuracyLow Range: ±0.03 mg/L ±3% of reading High Range: ±1 mg/L ±5% of reading
Silica MethodLow Range: Adaptation of the ASTM Manual of Water and Environmental Technology, D859, Heteropoly Molybdenum Blue method High Range: Adaptation of the USEPA Method 370.1 and Standard Method 4500-SiO2
SILVER
Silver Range0.000 to 1.000 mg/L (as Ag)
Silver Resolution0.001 mg/L
Silver Accuracy±0.020 mg/L ±5% of reading
Silver MethodAdaptation of the PAN method
SULPHATE
Sulphate Range0 to 150 mg/L (as SO42-)
Sulphate Resolution1 mg/L
Sulphate Accuracy±5 mg/L ±3% of reading
Sulphate MethodTurbidimetric - Sulphate is precipitated with barium chloride crystals
SURFACTANTS, ANIONIC
Range0.0 to 3.50 mg/L (as SDBS)
Resolution0.01mg/L
Accuracy±0.04 mg/L ±3% of reading
MethodMethylene blue
ZINC
Zinc Range0.00 to 3.00 mg/L (as Zn)
Zinc Resolution0.01 mg/L
Zinc Accuracy±0.03 mg/L ±3% of reading
Zinc MethodAdaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, Zincon method
REAGENTS
AlkalinityHI-93755-01 (100 tests); HI-93755-03 (300 tests)
AluminiumHI-93712-01 (100 tests); HI-93712-03 (300 tests)
Ammonia HR, MR & LRHR - HI-93733-01 (100 tests); HI-93733-03 (300 tests); MR - HI-93715-01 (100 tests); HI-93715-03 (300 tests); LR - HI-93700-01 (100 tests); HI-93700-03 (300 tests)
BromineHI-93716-01 (100 tests); HI-93716-03 (300 tests)
CalciumHI-937521-01 (50 tests); HI-937521-03 (150 tests)
Calcium and MagnesiumHI-93752-01 (50 x 2 tests) HI-93752-03 (150 x 2 tests)
ChlorideHI-93753-01 (100 tests); HI-93753-03 (300 Tests)
Chlorine DioxideReagents: HI-93738-01 (100 tests); HI-93738-03 (300 tests)
Chlorine - FreeFree Chlorine Reagents: HI-93701-01 (100 tests); HI-93701-03 (300 tests); HI-93701-F (liquid) (300 tests)
Chlorine UHRHI-95771-01 (100 tests); HI-95771-03 (300 tests)
Chlorine, Free and Total HRHI-93734-01 (100 tests); HI-93734-03 (300 tests)
Chlorine, Free ULRHI-95762-01 (100 tests); HI-95762-03 (300 tests)
Chlorine TotalHI-93711-01 (100 tests); HI-93711-03 (300 tests); HI-93711-T (300 tests liquid)
Chlorine, Total ULR: HI-95761-01 (100 tests); HI-95761-03 (300 tests)
Chromium VI – HR & LRHR - HI-93723-01 (100 tests); HI-93723-03 (300 tests)
LR - HI-93749-01 (100 tests); HI-93749-03 (300 tests)
Copper – HR, total & LRHR - HI-93702-01 (100 tests); HI-93702-03 (300 tests)
Total copper - HI-93702T - 01 (100 tests); HI-93702T-03 (300 tests)
LR - HI-95747-01 (100 tests); HI-95747-03 (300 tests)
CyanideHI-93714-01 (100 tests); HI-93714-03 (300 tests)
Cyanuric AcidHI-93722-01 (100 tests); HI-93722-03 (300 tests)
Detergents, AnionicHI-95769-01 (40 tests)
Fluoride – HR & LRHR - HI-93739-01 (100 tests); HI-93739-03 (300 tests)
LR - HI-93729-01 (100 tests); HI-93729-03 (300 tests)
Glycine PowderHI-93703-52 (100 tests)
Hardness, CalciumHI-93720-01 (100 tests); HI-93720-03 (300 tests)
Hardness, Magnesium and total hardnessHI-93719-01 (100 tests); HI-93719-03 (300 tests)
Hardness, total HR, MR, LR and Total LR+MR+LRHR - HI-93735-02 (100 tests); MR - HI-93735-01 (100 tests); LR - HI-93735-00 (100 tests); HR MR &LR - HI-93735-0 (100 x 3 tests)
HydrazineHI-93704-01 (100 tests); HI-93704-03 (300 tests)
IodineHI-93718-01 (100 tests); HI-93718-03 (300 tests)
Iron – HR & LRHR - HI-93721-01 (100 tests); HI-93721-03 (300 tests)
LR - HI-93746-01 (100 tests); HI-93746-03 (300 tests)
Manganese – HR & LRHR - HI-93709-01 (100 tests); HI-93709-03 (300 tests)
LR - HI-93748-01 (50 tests); HI-93748-03 (150 tests)
MagnesiumHI-937520-01 (50 tests); HI-937520-03 (150 tests)
MolybdenumHI-93730-01 (100 tests); HI-93730-03 (300 tests)
Nickel - HR & LRHR - HI-93726-01 (100 tests); HI-93726-03 (300 tests)
LR - HI-93740-01 (50 tests); HI-93740-03 (150 tests)
NitrateHI-93728-01 (100 tests); HI-93728-03 (300 tests)
Nitrite – HR & LRHR - HI-93708-01 (100 tests); HI-93708-03 (300 tests)
LR - HI-93707-01 (100 tests); HI-93707-03 (300 tests)
Oxygen, Dissolved (DO)HI-93732-01 (100 tests); HI-93732-03 (300 tests)
OzoneHI-93757-01 (100 tests); HI-93757-03 (300 tests)
pHHI-93710-01 (100 tests); HI-93710-03 (300 tests)
Phosphate - HR & LRHR - HI-93717-01 (100 tests); HI-93717-03 (300 tests)
LR - HI-93713-01 (100 tests); HI-93713-03 (300 tests)
PhosphorusHI-93706-01 (100 tests); HI-93706-03 (300 tests)
PotassiumHI-93750-01 (100 tests); HI-93750-03 (300 tests)
Silica – HR & LRHR - HI-96770-01 (100 tests); HI-96770-03 (300 tests)
LR - HI-93705-01 (100 tests); HI-93705-03 (300 tests)
SilverHI-93737-01 (50 tests); HI-93737-03 (150 tests)
SulphateHI-93751-01 (100 tests); HI-93751-03 (300 tests)