Hanna Instruments

Scientific Equipment

Safety Data Sheets

SDS search & quick download

Free Delivery

Most of mainland GB for orders over £50. Other charges may apply - see more

Hanna Care

We take the long view, looking after existing customers as much as delighting new ones!

Certificates of Analysis

Download a copy of the certificate of analysis for your solution

BL-983314-0 Resistivity Mini Controller

BL-983314-0 Resistivity Mini Controller

Hanna BL-983314-0 Resistivity Mini Controller

BL-983314-0 Resistivity Mini Controller

£288.20 (exc. VAT) £345.84 (inc. VAT)

The BL-983314 is a simple to operate resistivity controller designed for ultra pure water, reverse osmosis, and water conditioning applications. The BL-983314 resistivity controller is also ideal for continuous monitoring to process solutions. Setpoint and calibration are manually adjusted with a trimmer and the alarm relay allows for simple control.

BL-983314-0 Resistivity Mini Controller information

Hanna's extended description

About the BL-983314 Resistivity Mini Controllers

  • Fire retardant casing
  • Selectable overdose protection system
  • Splash-resistant cover

  • Readings are automatically temperature compensated with three different coefficients (β=2.4, 3.5 or 4.5 %/°C). The alarm contact can be used for connection to an alarm, pump, solenoid or dosing system.

    The relay contact is open when readings are higher than the setpoint, while for measurements lower than setpoint, the relay contact is closed. The hysteresis is typically 0.20 MΩ •cm from the setpoint.

    Measurements are displayed on the LCD and the multi-coloured LED continuously indicates if the controller is in measurement, dosing, or alarm mode. Users can choose automatic or manual dosing mode with a switch on the front panel. Manual control is particularly useful during maintenance operations, because it permits operators to enable or disable the dosing relay according to need.

    The overtime control system advises users when the relay is active for too long, helping to prevent over-dosage.

    Specifications

    Specifications list

    Specifications

    Range 0.00 to 19.90 MΩ•cm
    Resolution 0.10 MΩ•cm
    Accuracy (@25°C/77°F) ±2% F.S.
    Calibration factory calibrated
    Dosing Relay maximum 2A (fuse protected), 250 Vac, 30 Vdc contact closed when measure < setpoint
    Setpoint adjustable from 0 to 19.90 MΩ•cm
    Overtime adjustable, typically from 5 to approximately 30 minutes
    Temperature Compensation automatic and linear from 5 to 50°C (41 to 122°F) [β=2.4 ; 3.5 ; 4.5 %/°C selectable through jumper on the rear panel]
    Probe HI-3314 resistivity probe with 2 m (6.6’) cable (included)
    Power Supply BL983314-0: 12 VDC adapter (included); BL983314-1: 115/230 VAC; 50/60Hz
    Dimensions 83 x 53 x 99 mm (3.3 x 2.1 x 3.9’’)
    Weight BL983314-0: 200 g (7.1 oz.); BL983314-1: 300 g (10.6 oz.)
    Ordering Information BL983314-0 (12 VDC) and BL983314-1 ( 115/230V) are supplied with mounting brackets, transparent cover and instruction manual.